<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>37264</ErrorNumber>
<ErrorFamily>Head Interface</ErrorFamily>
<ErrorMessage>
HIF: ENQQ expected and not received
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The exposure head is in programming mode but the response to the programming
command is corrupted.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Check power supplies, cabling to the exposure head, etc.
</text></DescriptionText>
</OperatorAction>
<Availability>
<DescriptionText><text>
This error is reported.
</text></DescriptionText>
</Availability>
<Cause>
The exposure head power supply is faulty.
</Cause>
<Cause>
The cabling to the exposure head is faulty.
</Cause>
<LastModified>
30-Jul-2002
</LastModified>